Instrument features
1. High accuracy
Sample analysis of the whole high temperature heat, the analysis of the sample without loss;
Using gas chromatography detection, analysis methods in line with national standards, accurate measurement;
The use of electronic pressure flow control technology (EPC, EFC), respectively, control the carrier gas, air and hydrogen, to ensure high repeatability and high accuracy.
2. Qualitative and diverse
By increasing the flow path, a single cycle, can achieve non-methane total hydrocarbons, benzene, toluene, xylene at the same time detection.
3. stable and reliable
The whole line of electronic flow control (EPC), stable and reliable;
FID detector with automatic ignition and temperature judging function, FID flame automatically turn off after the hydrogen and air flow to ensure system security;
Core components are used well-known brand devices abroad, high reliability, long life
4. Real-time monitoring
Single cycle time is less than 90s, to ensure real-time monitoring.
5. highly intelligent
Regular automatic calibration and calibration;
Touch screen man-machine interface, simple operation, easy to use;
Fully automatic operation, unattended.
System Features
Using standard gas chromatography analysis method, accurate measurement.
Using the whole high temperature heating design, from sampling to analysis of the whole high temperature, to avoid the attachment of high boiling point VOCs to prevent corrosion of parts.
The system uses the main gas pumping, bypass analyzer intake of the sampling method to ensure that the system response time.
The system is filtered through three stages to ensure the life of the analyzer
Type : methane
Applications : environmental
Interface : Ethernet, RS-485
Other characteristics : online